FlashLink® USB Logger
Reusable Data Logger with External Probe
![]() |
![]() |
- Single channel, external probe for product core temperature
- User Programmable: manual or automatic start, sample interval, alarm limits, data tags
- Large, easy to read LCD display
- Trip history statistics available on LCD
- NIST traceable and CE Mark
- USB Connection

| Feature | Specification |
|---|---|
| Probe Measurement Range | -40°F to 185°F (-40°C to 85°C) |
| Probe Size: Penetration tip (20902) | 0.14in Dia x 3.98in L (3.5mm Dia x 101mm L), 3.28 ft (1m) Teflon cable |
| Blunt tip (20903) | 0.15in Dia x 0.98in L (3.8mm Dia x 25mm L), 3.28 ft (1m) Teflon cable |
| Temperature Resolution | 0.1°F/°C |
| Accuracy | ±1.0°F from 14°F to 77°F (±0.5°C from -10°C to 25°C); ±2.0°F above 77°F (±1.0°C above 25°C) ±2.0°F below 14°F (±1.0°C below -10°C) |
| Data Capacity | 3823 points |
| Data Download Time | 2 seconds typical for 3823 points |
| Programable Alarm Levels | High & Low levels over entire range |
| Battery | (CR2032) 3.6V Lithium/18 months |
| Dimensions/Body | 1.575" x 3.189" x 0.575" (40mm x 81mm x 13mm) |
| Casing | PC/ABS plastic |
| Certification | CE Mark, NIST traceable, RoHS Compliant |
| Display | LCD, Red LED Alarm |
| Waterproof Rating | IP67 with silicon cap on |
| Storage conditions | -4°F to 122°F (-20°C to 50°C) 0% to 95% RH |
| Minimum System Requirements | Windows PC with XP, or higher |
| Programming & Downloading | Direct USB connection to PC, FlashPDF Program Manager software (order PN 20412) or USB Program Manager software (PN 20238), and Adobe Reader 7.0 or greater |
20902;20903